The i6000sflex is anAC current probe utilising the Rogowski principle. It can be used to measurecurrents up to 6000A when used in conjunction with oscilloscopes, recorders ordata loggers. The flexible and lightweight measuring head allows quick and easyinstallation in hard to reach areas.
| ElectricalCharacteristics |
| CurrentRanges |
|
60 A / 600 A / 6000 A ACRMS |
|
| OutputSensitivity (AC coupled) |
|
| LoadImpedance |
|
| Accuracy(at +25°C) |
|
±1% of range (45 65 Hz) |
|
| Linearity(10% to 100% of range) |
|
| Noise |
|
8mV ACRMS (60 A) |
|
2 mV ACRMS (600/6000A) |
|
| FrequencyRange |
|
| PhaseError |
|
<±1º (45 65Hz), |
|
±10º (at 20 kHz) |
|
| PositionSensitivity (with cable> 25mm from the coupling) |
|
| Externalfield (with cable> 200mm from the head) |
|
| PowerSupply |
|
2 x AA MN 1500 LR6 alkaline |
|
400 Hours, low battery indicator or dedicated ext power supply |
|
| TemperatureCoefficient |
|
| WorkingVoltage (see Safety Standards section) |
|
| GeneralCharacteristics |
| Head Cablelength
(double insulated) |
|
610mm (24) I6000s Flex-24 915mm (36) I6000s Flex-36 |
|
| Head CableDiameter |
|
| Head BendRadius |
|
| Cablelength (head to electronics) |
|
| OutputConnection |
|
0.5 m cable terminated with safety BNC connector Supplied with 4mm safety plug adaptor |
|
| OperatingTemperature Range |
|
-20 to +90 °C (head) -20 to +85 °C (electronics) |
|
| StorageTemperature Range |
|
-40 to +105 °C (head) -20 to +85 °C (electronics) |
|
| OperatingHumidity |
|
15% to 85% (non condensing) |
|
| Weight |
|
180 g (head), 190 g (electronics) |
|
| SafetyStandards |
| BS EN61010 |
|
| BS EN61010 |
|
| BS EN61010 |
|
| 600VRMS,Category III, Pollution Degree 2 |
|
| Note |
|
Use of the probe on uninsulated conductors is limited to 600 V ACRMS or DC and frequencies below 1 kHz. |
|
| EMCStandards |
| BS EN 61326: |
1998 +A1, A2, & A3 |
|