| General Specifications |
| Maximum voltage applied to any terminal: |
|
| Storage Temperature |
|
-40 øC to 60 øC (-40 øFto 140 øF) |
|
| Operating Temperature |
|
-20 øC to 55 øC (-4 øFto 131 øF) |
|
| Temperature Coefficient |
|
0.05 x (specifiedaccuracy) per øC for temperatures 28 øC( 82 øF) |
|
| Relative Humidity |
|
0 % to 95 % @ 10 øC to 30øC (50 øF to 86 øF) |
|
0 % to 75 % @ 30 øC to 40øC (86 øF to 104 øF) |
|
0 % to 40 % @ 40 øC to 55øC (104 øF to 131 øF) |
|
| Vibration |
|
Random, 2 g, 5-500 Hz perMIL-PRF-28800F, Class 2 instrument |
|
| Shock |
|
1 meter drop per IEC61010-1 2nd Edition (1 meter drop test, six sides, oak floor) |
|
| Electromagnetic compatibility: |
|
In an RF field of 3 V/M,accuracy = specified accuracy (EN 61326-1:1997) |
|
| Safety |
|
Complies with ANSI/ISA82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN61010-1 2nd Edition for measurement category IV 600 V (CAT IV) |
|
| Certifications: |
|
CSA per standard CSA/CANC22.2 No. 61010.1-04; TUV per standard IEC/EN 61010-1 2ndEdition |
|
| Batteries |
|
Four AA batteries (NEDA15A or IEC LR6) |
|
| Battery Life |
| Insulation testuse: |
Tester can perform atleast 1000 insulation tests with fresh alkaline batteries atroom temperature. These are standard tests of 1000 V into 1 MΩwith a duty cycle of 5 seconds on and 25 seconds off. |
| Resistancemeasurements: |
Tester can perform atleast 2500 earth bond resistance measurements with freshalkaline batteries at room temperature. These are standardtests of 1 Ω with a duty cycle of 5 seconds on and 25seconds off. |
|
| Size |
|
5.0 cm H x 10.0 cm W x20.3 cm L (1.97 in H x 3.94 in W x 8.00 in L) |
|
| Weight |
|
| IP Rating |
|
| Altitude |
| Operating: |
2000 m CAT IV 600 V, 3000m CAT III 600 V |
| Non-operating(storage): |
12,000 m |
|
| Over-range capability: |
|
| Included accessories: |
|
TL224 Test Leads, TP74Test Probes, clips PN 1958654 (red) and PN 1958646 (black),holster and remote probe |
|