The Fluke 709 mA loop calibrator isdesigned to save time and produce high-quality results. This tool reduces thetime it takes to measure, or source, voltage or current and power up a loop
| ModelName | Description |
| Fluke709 |
Fluke 709H Precision Current Loop Calibratorwith HART Communications/Diagnostics including:
|
HART communication
The Fluke 709H offers a built-in HART modem for communication capability to perform the following commands:
- Read Message
- Read Tag, Descriptor, Cal Date
- Read Sensor PV Info
- Read PV Output Info
- Read Long Tag
- Write PV Ranges (Upper and Lower)
- Enter/Exit Fixed Current Mode
- Set Zero Offset
- Trim DAC Zero
- Trim DAC Gain
The Fluke 709H also offers ability to store up to twenty HART device configuration files for uploading via 709H/Track software. Configurations can be stored as .csv or .txt files. This allows the end user to document the entire plant for HART devices without spending thousands of dollars for plant asset management software
| DetailedSpecifications | |
| Functions | mA source, mA simulate, mA read, mAread/loop power, and volts read. |
| Ranges | mA (0 to 24mA) and Volts (0 to 30VDC) |
| Resolution | 1 µA on mA ranges and 1 mV on voltage range |
| Accuracy | 0.01% +/- 2 counts, all ranges (@23° +/- 5°C) |
| Operating TemperatureRange | -10 °C to 55 °C(14 °F to 131°F) |
| Humidity Range | 10 to 95% non-condensing |
| Stability | 20 ppm of F.S. /°C from -10°C to 18°C and28°C to 55°C |
| Display | 128 x 64 pixels, LCD Graphic w/backlight,.34 high digits |
| Power | 6 AAA alkaline |
| Battery Life | 40 hours typical |
| Loop Compliance Voltage | 24VDC @ 20mA |
| Over-Voltage Protection | 240 VAC |
| Overload CurrentProtection | 28 mA DC |
| EMC | EN61326 Annex A (Portable Instruments) |
| Dimensions (L x W x D) | 15 cm x 9 cm x 3 cm(6 x 3.6 x 1.3) |
| Weight | 0.3 kg(9.5 ounces) |
| Included Accessories | NIST traceable calibration certificate withdata, batteries, test leads, and manual |
| Warranty | Three year |